Free Shipping for all Order above Rs.4999/-
Showing the single result
-
Patient Monitor HD-11V
₹24,999.00 incl. GSTFeature
- 12.1 inch, 800*600/1024×768 dpi, color TFT display:
- Suitable for OT and ICU:
- Separated module: Main board+Parameter module(ECG, NIBP, SpO2);
- Display brightness adjustable;
- Real time S-T segment analysis, and ARR analysis;
- SpO2 Pitch Tone Variation and Drug Dose Calculation;
- Parameter waveform and characters color selectable
- Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic
- Standby mode for power saving and suspend monitoring.
- 7-lead ECG waveforms display in phase,
- Capture dynamic waveforms and history waveforms review,
- VGA output:
- Built-in rechargeable lithium battery 4 hours, and battery volume checking, Efficient resistance to interference of defibrillator and HF knife;
- SD card memory and USB data output:
- TCP/IP networking platform enable abundant future upgrade& CMS;
- Complete and flexible mounting solution for transportation and
- In-patient monitoring.
SPECIFICATIONS:
ECG:
- Lead mode: 5-lead(R,L,F,N,C)
- Lead selection: I,II,III,avR, avl,avF,V
- Waveform: 3 and 7 channel selectable
- Gain selection: 0.5mm/mv,1mm/mv,2mm/mv
- Sweep speed: 12.5mm/s;25mm/s;50mm/s
Heart rate range:
Adult: 15~300bpm;
Neonate/Pediatric:15-350bpm
- Accuracy: +1bpm or +1%, whichecer is greater
- Resolution: 1bpm
- Filter: surgery mode:1~20Hz
- monitor model: 0.5~40Hz
- Diagnostic mode: 0.05~130Hz
- Scaling signal: 1mv,+3%
- Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
- Alarm range: 15~350bpm
SPO2
- Measurement range: 0~100%
- Resolution: 1%
- Accuracy: +2%(70-100%);0-69% unspecified
- Alarm range: 0~100%
- Pluse Rate: range:20~300bpm
- Resolution: 1bpm
- Error: +1bpm or +2%, whichever is greater
RESP:
- Method: Thoracic impedance
- Measurement range: Adult:7~120rpm;
- Neonate/Pediatric:
- 7~150rpm
- Apnea alarm: YES,10~40s
- Resolution: 1rpm
- Accuracy: +2rpm
TEMP:
- Compatible probe: YSI or CYF
- Measurement range: 5~50°C
- Resolution: 0.1°C
- Accuracy: +0.1°C
- Refreshing time: about 1
- Average measuring time: <10s
NIBP:
- Method: Digital Automatic oscillometric
- Operation mode: Manual/Automatic/continuous
- Auto measurement time: Adjustable(1~480min)
- Measurement Unit: mmHg/Kpa selectable
- Measurement types: Systolic, Diastolic, Mean
Measurement tange:
- Range of Systolic pressure:
- Adult: 40-270mmHg
- Pediatric: 40~220mmHg
- Neonate: 40~135mmHg
- Range of mean pressure:
- Adult: 20~235mmHg
- Pediatric: 20-165mmHg
- Neonate: 20~110mmHg
Range of diastolic pressure:
- Adult: 10~215mmHg
- Pediatric: 10~150mmHg
- Neonate:10~100mmHg
Over-pressure protection:
- Double safety protection
- Resolution: 1 mmHg
- Alarm: Systolic.Diastolic, Mean
- IBP (Optional)
- IBP Waveform: 2-Channel
- Max 8 channels
- Measurement Range: -50-350mmHg
- Resolution: 1mmHg:
- Unit: mmHg, kPa
Accuracy:
- Static:±1mmHg or 2%, whichever is greater
- (exclusive of transducer)
- Dynamc: ±4mmHg or 4%, whichever is greater
- Sensitivity of transducer: 5 uV/V/mmHg, ±2%
- Impedance of transducer:300-3000 Ohm
Transducer sites:
- ART, PA, LAP, RAP,CVP, ICP,P1,P2
- Selection of measurement range:
- ART: 0~+350mmHg
- PA:-10~+120mmHg
- CVP/RAP/LAP/ICP:-10~+40mmHg
- P1/P2:-50+360mmHg
EtCO2( Sidertream CO2, Optional)
- Measurement range: 0~99mmHg
- Accuracy: +2mmHg (0~40mmHg)
- Sampling Range: 100ml/min
- Sampling rate accuracy: 15%
- Respiration rate: 0~120rmp
- Respiration accuracy: +2rmp (0~70rmp)
- +5rmp (>70rmp)
- Respiration time: 240msec (10% to 90%)
- Delay time: <25
EtCO2 (Mainstreaam CO2, Optional)
- Method: Infrared Spectrum
- Range: 0.0-10% (0~76%)
- Resolution: 1mmHg (0.1%)
- Accuracy: <5%(±4.0 mmHg)
- Or 10% (of Readings)
Suntech(Suntech):
- Technique: Oscillometric
- Patient Applications: Adult Pediatric, Neonate
- Systolic Range: 40-265mmHh
- Diatolic Range: 20-200mmHh
- MAP Range: 27-222mmHh
- Pulse Range: 25-300BPM
- Transducer Accuracy: ±3mHg up to full range
- Operation Conditions:
- 0°C to 50°C, 15% to 95% non-condensing RH
- Storage Conditions:
- -20°C to 65 °C, 15%to 95% non-condensing RH
Recorder (Optional)
- Build-in, thermal array
- Plethysmogram caveform: 2 channels
- Record mode: manual, on alarm, time-defined
- Recording width: 50mm
- Printing speed: 50mm/s
- Recording type: Frozen waveform record
- NIBP recall record
- Trend table record
- Alarm record
- Fixed-time record