• Patient Monitor HD-11V

    24,999.00 incl. GST

    Feature

    • 12.1 inch, 800*600/1024×768 dpi, color TFT display:
    • Suitable for OT and ICU:
    • Separated module: Main board+Parameter module(ECG, NIBP, SpO2);
    • Display brightness adjustable;
    • Real time S-T segment analysis, and ARR analysis;
    • SpO2 Pitch Tone Variation and Drug Dose Calculation;
    • Parameter waveform and characters color selectable
    • Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic
    • Standby mode for power saving and suspend monitoring.
    • 7-lead ECG waveforms display in phase,
    • Capture dynamic waveforms and history waveforms review,
    • VGA output:
    • Built-in rechargeable lithium battery 4 hours, and battery volume checking, Efficient resistance to interference of defibrillator and HF knife;
    • SD card memory and USB data output:
    • TCP/IP networking platform enable abundant future upgrade& CMS;
    • Complete and flexible mounting solution for transportation and
    • In-patient monitoring.

    SPECIFICATIONS:

    ECG:

    • Lead mode: 5-lead(R,L,F,N,C)
    • Lead selection: I,II,III,avR, avl,avF,V
    • Waveform: 3 and 7 channel selectable
    • Gain selection: 0.5mm/mv,1mm/mv,2mm/mv
    • Sweep speed: 12.5mm/s;25mm/s;50mm/s

    Heart rate range:

    Adult: 15~300bpm;

    Neonate/Pediatric:15-350bpm

    • Accuracy: +1bpm or +1%, whichecer is greater
    • Resolution: 1bpm
    • Filter: surgery mode:1~20Hz
    • monitor model: 0.5~40Hz
    • Diagnostic mode: 0.05~130Hz
    • Scaling signal: 1mv,+3%
    • Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
    • Alarm range: 15~350bpm

    SPO2

    • Measurement range: 0~100%
    • Resolution: 1%
    • Accuracy: +2%(70-100%);0-69% unspecified
    • Alarm range: 0~100%
    • Pluse Rate: range:20~300bpm
    • Resolution: 1bpm
    • Error: +1bpm or +2%, whichever is greater

    RESP:

    • Method: Thoracic impedance
    • Measurement range: Adult:7~120rpm;
    • Neonate/Pediatric:
    • 7~150rpm
    • Apnea alarm: YES,10~40s
    • Resolution: 1rpm
    • Accuracy: +2rpm

    TEMP:

    • Compatible probe: YSI or CYF
    • Measurement range: 5~50°C
    • Resolution: 0.1°C
    • Accuracy: +0.1°C
    • Refreshing time: about 1
    • Average measuring time: <10s

    NIBP:

    • Method: Digital Automatic oscillometric
    • Operation mode: Manual/Automatic/continuous
    • Auto measurement time: Adjustable(1~480min)
    • Measurement Unit: mmHg/Kpa selectable
    • Measurement types: Systolic, Diastolic, Mean

    Measurement tange:

    • Range of Systolic pressure:
    • Adult: 40-270mmHg
    • Pediatric: 40~220mmHg
    • Neonate: 40~135mmHg
    • Range of mean pressure:
    • Adult: 20~235mmHg
    • Pediatric: 20-165mmHg
    • Neonate: 20~110mmHg

    Range of diastolic pressure:

    • Adult: 10~215mmHg
    • Pediatric: 10~150mmHg
    • Neonate:10~100mmHg

    Over-pressure protection:

    • Double safety protection
    • Resolution: 1 mmHg
    • Alarm: Systolic.Diastolic, Mean
    • IBP (Optional)
    • IBP Waveform: 2-Channel
    • Max 8 channels
    • Measurement Range: -50-350mmHg
    • Resolution: 1mmHg:
    • Unit: mmHg, kPa

    Accuracy:

    • Static:±1mmHg or 2%, whichever is greater
    • (exclusive of transducer)
    • Dynamc: ±4mmHg or 4%, whichever is greater
    • Sensitivity of transducer: 5 uV/V/mmHg, ±2%
    • Impedance of transducer:300-3000 Ohm

    Transducer sites:

    • ART, PA, LAP, RAP,CVP, ICP,P1,P2
    • Selection of measurement range:
    • ART: 0~+350mmHg
    • PA:-10~+120mmHg
    • CVP/RAP/LAP/ICP:-10~+40mmHg
    • P1/P2:-50+360mmHg

    EtCO2( Sidertream CO2, Optional)

    • Measurement range: 0~99mmHg
    • Accuracy: +2mmHg (0~40mmHg)
    • Sampling Range: 100ml/min
    • Sampling rate accuracy: 15%
    • Respiration rate: 0~120rmp
    • Respiration accuracy: +2rmp (0~70rmp)
    • +5rmp (>70rmp)
    • Respiration time: 240msec (10% to 90%)
    • Delay time: <25

    EtCO2 (Mainstreaam CO2, Optional)

    • Method: Infrared Spectrum
    • Range: 0.0-10% (0~76%)
    • Resolution: 1mmHg (0.1%)
    • Accuracy: <5%(±4.0 mmHg)
    • Or 10% (of Readings)

    Suntech(Suntech):

    • Technique: Oscillometric
    • Patient Applications: Adult Pediatric, Neonate
    • Systolic Range: 40-265mmHh
    • Diatolic Range: 20-200mmHh
    • MAP Range: 27-222mmHh
    • Pulse Range: 25-300BPM
    • Transducer Accuracy: ±3mHg up to full range
    • Operation Conditions:
    • 0°C to 50°C, 15% to 95% non-condensing RH
    • Storage Conditions:
    • -20°C to 65 °C, 15%to 95% non-condensing RH

    Recorder (Optional)

    • Build-in, thermal array
    • Plethysmogram caveform: 2 channels
    • Record mode: manual, on alarm, time-defined
    • Recording width: 50mm
    • Printing speed: 50mm/s
    • Recording type: Frozen waveform record
    • NIBP recall record
    • Trend table record
    • Alarm record
    • Fixed-time record
  • Patient Monitor HD-PM12T

    26,499.00 incl. GST

    Display Size -12.1 Inch

    Type of display – LCD

    Brand – Dr Diaz

    Model – PM 12T

    Parameters Measured – SPO2, Temperature, RESP, ETCO2, NIBP, ECG, IBP

    Suitable For – Pediatric, Adult, Neonatal

  • Patient Monitor HD-PM15B

    31,999.00 incl. GST

    Features:

    • 15.6 inch large screen provides bright and clear view
    • Multi-language operation system
    • Sync display of 7-lead ECG
    • VGA output for external display
    • USB interfaces support easy software upgrade
    • Three working modes: Monitoring, Surgery and Diagnosis.
    • Central monitoring system by wired mode.

      Standard Parameters:
    • ECG, RESP, TEMP, NIBP, SpO2, PR
    • Optional Parameters:

      • Dual IBP
      • Etco2
      • 12-lead ECG
      • Recorder